Breast Flat Panel Detector MSLFP30 for sale|Medsinglong
1.
Introduction
The
MSLFP30
is a digital FPD detector with the patented Amorphous Silicon TFT sensor and needles directly deposited CsI: Tl scintillator technique of iRay technology.
It
is designed with the compact outline in order to retrofit the analog mammography system to digital.
MSLFP30
has a quite large active area of 24x30cm with 85μm pixel pitch. The employed CsI
(Cesium
Iodide)
scintillator
makes
the
MSLFP30
achieve
a
special
high
DQE
and
MTF.
The detector provides AED (Auto Exposure Detection) function via the specialized sensor to detect the X-ray for activing the FPD, and is also compatible with the existed AEC function of the analog mammography system.
MSLFP30
supports FFDM
applications.
Sensor technology | Amorphous silicon TFT/PD matrix panel |
Scintillator | CsI (TI), needles directly deposited on the a-Si TFT panel |
Pixel pitch | 85µm |
Effective active area | 235mm x 290mm |
Effective pixel matrix | > 2762 x 3408 |
AD conversion | 16bit |
Charge amplifier | Programmable gain amplifier |
Energy range | 25kV-40kV |
DataControl interface | GigE |
Trigger mode | Software, AED |
AED function | Yes |
Frame rate | 2fps@Free run |
Cycle time | Typical 15s |
- Image Quality
Limited spatial resolution | 6 lp/mm |
Saturation dose* | Typical 2500μGy @ 7.2pC |
Dynamic range* | >76dB @ 7.2pC |
QNED (measured according to European Guidelines) | FFDM: ≤20µGy @ Binning 1C |
QNLD (measured according to European Guidelines) | FFDM: ≤80µGy @ Binning 1C |
Linearity* | ≥ 0.999 |
MTF* @ RQM1/Binning1C (measured according to IEC62220-1-2) | |
DQE* @ RQM1/Binning1C (measured according to IEC62220-1-2) | |
Lag (measured according to IEC62220-1-2) | < 1.5% (first frame) |
Memory effect | <0.3%, after 60s |
(measured according to IEC62220-1-2) | |
CNR @ 45mm PMMA | ≥ 12.5 |
Image receptor homogeneity (measured according to European Guidelines) | <10% |
ACR phantom @ clinical condition | ≥ 4 fibers ≥ 4 masses ≥ 3 speck groups |
Missed tissue at the chest wall side | < 3 mm |
*: The Radiation quality is W/Rh (50µm), according to
IEC62220-1-2:2007
.
- Mechanical
Weight | ≤1.3kg |
Housing material | Aluminum with carbon fiber laminate entrance window |
Size (mm3) | 327 (L) × 254.5 (W) × 14 (H) |
Outline |
- Communications
Data interface | GigE |
X-ray synchronization interface | AED |
Software | SDK, supporting the 32 and 64 Windows® OS |
Temperature | 10 – 40 °C (operating), -10 – 55 °C (storage) |
Humidity | 10 – 90% RH (operating, non-condensing) 10 – 95%RH (storage) |
Pressure | 700 – 1060mbar (operating and storage) |
Vibration | IEC 60068-2-6 (10-200 Hz, 5 g) |
Shock | IEC 60068-2-29 (16ms, 10 g) |
Life time | 5 years |
Life time dose | Typical 300Gy |
IBFR | ≤ 2.5% |
CFR36 | ≤ 7.5% |
- Power
Supply | 100-240V AC and mains frequency of 50Hz and 60Hz, both +/- 2%. |
Input | 24V DC |
Dissipation | Around 18W |
Cooling | Passive |
- Regulatory
General regulatory | CE, RoHs |
Risk management | ISO 14971: 2007 |
Safety and essential perfor mance | IEC60601-1:2005/EN60601- 1:2006+AC:2010/IEC 60601-1:2005+ Amendment 1:2012/EN 60601-1:2006+ Amendment 1:2013 IEC 60601-1-2:2014 |
Radiation protection | IEC 60601-1-3:2008 |
Characteristics of digital X-ray imaging devices | IEC 62220-1-2:2007 |