Reliable AED function Mammography Detector FPD Digital X-ray Detector Digital Flat Panel Detector

Reliable AED function Mammography Detector FPD Digital X-ray Detector Digital Flat Panel Detector

10 La gente viendo este producto ahora!
  • Retiro en Casa Matriz

Haz tu pedido y retira directamente

1-2 Días

Gratuito

  • Envío a Domicilio
Antes de comprar verifica el día en que despachamos a tu comuna

2-3 Días

Costo Variable

  • Certificado de Calidad
  • Empaque y Traslado Seguro

Prefiera Tarjetas Débito o Crédito:

Breast Flat Panel Detector MSLFP30 for sale|Medsinglong

 

1.

Introduction

 

The 

MSLFP30

 is a digital FPD detector with the patented Amorphous Silicon TFT sensor and needles directly deposited CsI: Tl scintillator technique of iRay technology. 

It 

is designed with the compact outline in order to retrofit the analog mammography system to digital. 

MSLFP30

 has a quite large active area of 24x30cm with 85μm pixel pitch. The employed CsI

(Cesium

Iodide)

scintillator

makes

the

MSLFP30

achieve

a

special

high

DQE

and

MTF. 

The detector provides AED (Auto Exposure Detection) function via the specialized sensor to detect the X-ray for activing the FPD, and is also compatible with the existed AEC function of the analog mammography system. 

MSLFP30

 supports FFDM

applications.

 

  1.  

 

 

Sensor technology Amorphous silicon TFT/PD matrix panel
Scintillator CsI (TI), needles directly deposited on the a-Si TFT panel
Pixel pitch 85µm
Effective active area 235mm x 290mm
Effective pixel matrix > 2762 x 3408
AD conversion 16bit
Charge amplifier Programmable gain amplifier
Energy range 25kV-40kV
DataControl interface GigE
Trigger mode Software, AED
AED function Yes
Frame rate 2fps@Free run
Cycle time Typical 15s

 

 

  1. Image Quality

 

 

Limited spatial resolution 6 lp/mm
Saturation dose* Typical 2500μGy @ 7.2pC
Dynamic range* >76dB @ 7.2pC
QNED (measured according to European Guidelines) FFDM: ≤20µGy @ Binning 1C
QNLD (measured according to European Guidelines) FFDM: ≤80µGy @ Binning 1C
Linearity* ≥ 0.999
MTF* @ RQM1/Binning1C (measured according to IEC62220-1-2)        
DQE* @ RQM1/Binning1C (measured according to IEC62220-1-2)          
Lag (measured according to IEC62220-1-2) < 1.5% (first frame)
Memory effect <0.3%, after 60s

 

 

(measured according to IEC62220-1-2)
CNR @ 45mm PMMA ≥ 12.5
Image receptor homogeneity (measured according to European Guidelines) <10%
ACR phantom @ clinical condition ≥ 4 fibers ≥ 4 masses ≥ 3 speck groups
Missed tissue at the chest wall side < 3 mm

*: The Radiation quality is W/Rh (50µm), according to 

IEC62220-1-2:2007

.

 

  1. Mechanical

 

 

Weight ≤1.3kg
Housing material Aluminum with carbon fiber laminate entrance window
Size (mm3) 327 (L) × 254.5 (W) × 14 (H)
Outline

 

 

  1. Communications

 

 

Data interface GigE
X-ray synchronization interface AED
Software SDK, supporting the 32 and 64 Windows® OS

 

 

 

  1.  

 

 

Temperature 10 – 40 °C (operating), -10 – 55 °C (storage)
Humidity 10 – 90% RH (operating, non-condensing) 10 – 95%RH (storage)
Pressure 700 – 1060mbar (operating and storage)
Vibration IEC 60068-2-6 (10-200 Hz, 5 g)
Shock IEC 60068-2-29 (16ms, 10 g)
Life time 5 years
Life time dose Typical 300Gy
IBFR ≤ 2.5%
CFR36 ≤ 7.5%

 

 

 

  1. Power

 

 

Supply 100-240V AC and mains frequency of 50Hz and 60Hz, both +/- 2%.
Input 24V DC

 

 

Dissipation Around 18W
Cooling Passive

 

 

 

  1. Regulatory

 

 

General regulatory CE, RoHs
Risk management ISO 14971: 2007
Safety and essential perfor mance IEC60601-1:2005/EN60601- 1:2006+AC:2010/IEC 60601-1:2005+ Amendment 1:2012/EN 60601-1:2006+ Amendment 1:2013 IEC 60601-1-2:2014
Radiation protection IEC 60601-1-3:2008
Characteristics of digital X-ray imaging devices IEC 62220-1-2:2007

 

 